Tech Specs

Tech Specs

Defibrillator

TYPE
Semi-automatic external
defibrillator

MODELS
DDU-100A, DDU-100E

WAVEFORM
Biphasic Truncated Exponential
(Impedance compensated)

ENERGY
Adult: 150 Joules
Pediatric: 50 Joules
(nominal into 50 ohm load)

CHARGE TIME**

DBP-2800: Less than 6 seconds
DBP-1400: Less than 9 seconds

**Typical, with new battery at 25º C

VOICE PROMPTS
Extensive voice prompts
guide user through operation
of the unit and CPR

CPR PACING
Metronome

CONTROLS
– Lighted On/Off button
– Lighted Shock button

INDICATORS
– “check pads”
– “do not touch patient”
– “analyzing”
– AED status LED



Self Tests

AUTOMATIC
Automatic daily, weekly and
monthly and quarterly circuitry tests

BATTERY INSERTION
System integrity test on battery
insertion

PAD PRESENCE
Pads preconnected tested daily

USER-INITIATED
Unit and battery pack system
test may also be initiated by the
user

STATUS INDICATION
Visual and audible indication of
unit status


Patient Analysis System

PATIENT ANALYSIS
Automatically evaluates patient impedance for proper pad contact. Monitors signal quality and analyzes patient ECG for shockable/non-shockable rhythms

SENSITIVITY/SPECIFICITY
Meets AAMI-DF-39 specifications and AHA recommendations



Defibrillation / Monitoring Pads

MODEL
Adult – DDP-100
Child/Infant – DDP-200P

TYPE
Pre-connected, single-use, non-polarized, disposable, self-adhesive electrodes with cable and connector

SURFACE AREA
103 cm2 (nominal, each pad)
50 cm2 (nominal, each pad)

PAD PLACEMENT
Adult – Anterior/Anterior
Child/Infant – Anterior/Posterior

CABLE LENGTH (typical)
48 in (122 cm)


Download Tech Specs

Environmental

TEMPERATURE
Operating: 0 to 50°C (32 to 122°F)
Standby: 0 to 50°C (32 to 122°F)
(with battery installed)

RELATIVE HUMIDITY
Operating / Standby: 5% – 95%
(non-condensing)

ALTITUDE
-500 to 15,000 ft (-150 to 4500 m) per MIL-STD-810F 500.4 Procedure II

VIBRATION
Ground (MIL-STD-810F 514.5 Category 20)

Helicopter (RTCA/DO-160D, Section 8.8.2, Cat R, Zone 2, Curve G)

Jet Aircraft (RTCA/DO-160D, Section 8, Cat H, Zone 2, Curves B & R)

SHOCK / DROP ABUSE TOLERANCE
MIL-STD-810F 516.5 Procedure IV (1 meter, any edge, corner, or surface, in standby mode)

SEALING / WATER RESISTANCE
IEC60529 class IP54; Splash Proof, Dust Protected (Battery Pack installed)

ESD
EN61000-4-2:2009 (open air up to 15kV or direct contact up to 8kV)

EMC (Emission)
EN60601-1-2:2001 +A1:2006, limits EN55011:1998 +A1:1999 +A2:2002 Group 1 Level B, method

EMC (Immunity)
EN60601-2-4:2003, limits EN 61000-4-3:2002 Level 3 (10V/m), method


Battery Pack

TYPE
– Lithium/Manganese Dioxide
– Disposable, recyclable, non rechargeable

LOW BATTERY INDICATORS
– Visible
– Audible

MODEL DBP-2800

POWER
15V, 2800 mAh

CAPACITY **
– 300 shocks or 16 hours continuous operation

STANDBY-LIFE**
– 7 years

MODEL DBP-1400

POWER
15V, 1400 mAh

CAPACITY**
– 125 shocks or 8 hours
continuous operation

STANDBY-LIFE**
– 5 years

**Typical, with new battery at 25º C

 



Event Documentation

INTERNAL EVENT RECORD
Critical ECG segments and rescue event parameters are recorded and can be downloaded to a removable data card

PC-BASED EVENT REVIEW
ECG with event tag display, and audio playback when available

REMOVABLE STORAGE
(optional) Up to 12 hours of ECG and event data storage (no audio option) or up to 2 hours of audio, ECG and event storage (audio option) on a removable data card. Actual length of storage is dependent on card capacity.



SIZE

8.5 x 11.8 x 2.7 inches
(22 x 30 x 7 cm)

WEIGHT (Approximate)
With DBP-1400: 4.2 lbs (1.9 kg)
With DBP-2800: 4.4 lbs (2.0 kg)