Tech Specs

Tech Specs

Automated Chest Compression Device

TYPE
Automated Chest Compression Device

MODEL
RMU-1000


Compressions

COMPRESSION MODES

  • Continuous Compressions
  • Compressions with Breathing (30:2, 30 compressions with 3-second pause for ventilation) factory default; future protocols via field updates

COMPRESSION DEPTH
2.1 inches ±0.1 inches (5.3 cm ±0.3 cm) from start position (nominal patient)

COMPRESSION FREQUENCY
101 ±1 compressions per minute

COMPRESSION DUTY CYCLE
50% ±5%


Battery Pack

MODEL 
RBP-1000

TYPE
18.5V, 5300 mAh, Lithium-ion
Rechargeable, recyclable

OPERATION TIME 
1 hour (nominal patient) *

CHARGE TIME *
Less than 3 hours in ACC
Less than 2 hours in external battery pack charging station

USEFUL LIFE
Recommended to replace battery pack every 3 years or if battery pack indicator displays a replace battery pack condition (~300 charge/discharge cycles **)

OPERATING TEMPERATURE
0 to 40º C (32 to 104º F) ambient

CHARGING TEMPERATURE
0 to 40º C (32 to 104º F) ambient

STORAGE TEMPERATURE
0 to 40°C (32 to 104°F)
-20 to 60°C (-4 to 140°F) short-term <1 month 

SEALING / WATER RESISTANCE
IEC 60529 class IP44

 

 *  Typical, with new battery at 25º C

** One charge/discharge cycle is defined as charging and discharging the full capacity of the Battery Pack

 


Download Tech Specs

 

 

 

 

 

 

 

† Specifications subject to change without notice

Environmental

OPERATING / MAINTENANCE TEMPERATURE
0 to 40°C (32 to 104°F)

STANDBY / STORAGE / TRANSPORT TEMPERATURE
-20 to 70°C (-4 to 158°F)

HUMIDITY
5% – 95% (non-condensing)

VIBRATION
MIL-STD-810G 514.6 Category 20 (Ground)

SEALING / WATER RESISTANCE
IEC60529 class IP43 (battery pack installed)

ELECTROMAGNETIC COMPATIBILITY (EMISSIONS & IMMUNITY)
IEC 60601-1-2

DESIGN STANDARDS
Meets applicable requirements of:

  • IEC 60601-1
  • UL 60601-1
  • CAN/CSA C22.2 No.60601-1
  • IEC 60601-1-2

DEVICE CLASSIFICATION
Internally powered Class II (with external power source)


Physical

SIZE (assembled)
23.5 x 20.75 x 9 inches (59.7 x 52.7 x 22.9 cm)

SIZE (in carrying case)
20 x 20 x 10 inches (50.8 x 50.8 x 25.4 cm)

WEIGHT (with battery pack)
15.9 lbs. (7.1 kg)

ADULT PATIENT RANGES
Adult patients that fit into the ACC:

Chest width: 18 inches (45.7 cm) max.

Chest height: 6.5 to 11.8 inches (16.5 to 30 cm)

Use of the RMU-1000 is not restricted by patient weight.


AC Power Adapter

MODEL  
RPM-1000

RATED OUTPUT
24.0V ±5% @ 4.2A

INPUT VOLTAGE
85 to 264VAC (100 to 240VAC nominal) 

INPUT FREQUENCY
47 to 63Hz

INPUT CURRENT
< 2.3A rms

OPERATING TEMPERATURE
0 to 40º C (32 to 104º F) full load

STORAGE TEMPERATURE
-40 to 85º C (-40 to 185º F)

ELECTROMAGNETIC COMPATIBILITY (EMISSIONS & IMMUNITY)
IEC 60601-1-2

 

 